Measurement Uncertainty Calculator
Combine Type A (repeatability) and Type B (resolution) uncertainties.
About this tool
Enter the standard deviation of repeated readings, the number of readings and the instrument resolution. SD 0.1 over 4 readings with 0.01 resolution gives a combined standard uncertainty of 0.050083.
How to use
How it works
Type A uncertainty comes from repeated readings: s ÷ √n. Type B uncertainty from a digital resolution step is resolution ÷ √12, assuming a rectangular distribution. The two are combined as the square root of the sum of squares.
u_A = s ÷ √n u_B = resolution ÷ √12 u_c = √(u_A² + u_B²)
Worked example
A coarse instrument
- Input
- s = 0.05, n = 10, resolution = 0.1
- Output
- 0.032914
- Walkthrough
- u_A = 0.05 ÷ √10 = 0.015811; u_B = 0.1 ÷ √12 = 0.028868. Then √(0.015811² + 0.028868²) = 0.032914.
Limitations
- Covers repeatability and resolution only. Calibration and environmental terms must be added separately.
- It gives the standard uncertainty (1σ), not an expanded interval; multiply by a coverage factor such as 2 for about 95%.
Questions
What is the uncertainty for SD 0.1, 4 readings and 0.01 resolution?
What if the resolution is 0?
What is the SD 0.05, 10 readings, 0.1 resolution case?
Features
- Type A and Type B
- Combined standard uncertainty
- Runs in your browser
Related resources
Guides and videos from around the web—opens on YouTube or the publisher’s site.
Introduction to scientific notation
Khan Academy
Scientific notation for calculator and number tools.
Watch on YouTubeFinding a percentage
Khan Academy
Percentage basics used across many calculator pages.
Watch on YouTube
- Percentages
Math is Fun
Percent of / percent change patterns for everyday calculators.
- Data Index
Math is Fun
Stats and probability explainers for data tools.